Design of portable α - β - γ spectrometer
DOI:
Author:
Affiliation:

Clc Number:

TL817+.2

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
    Abstract:

    The existing portable spectrometer cannot measure the α-β-γ ray at the same time, which makes the evaluation and processing of the α-β-γ mixed scene inconvenient, so a portable α-β-γ spectrometer is developed to solve the problem. The spectrometer uses a PIPS detector and a LaBr3(Ce) detector to measure α-β particles and γ-rays, respectively, to obtain a high-resolution energy spectrum. The main properties of the spectrometer such as α energy resolution, α/β surface emissivity response, γ energy resolution and temperature stability were tested. The α energy resolution is less than 140 keV (5.486 meV, 0.1 atm), surface emissivity response of α particles is 67.34% (241Am), surface emissivity response of β particles is 19.13%(90Sr-90Y), and γ energy resolution is less than 3.4% (662 keV). Therefore, the spectrometer can simultaneously measure the α-β-γ rays and obtain the high resolution energy spectrum, which is helpful for the related personnel to make rapid and accurate evaluation and treatment of the measurement site.

    Reference
    Related
    Cited by
Get Citation

Cite this article as: LIU Yi-Wen, ZHAO Chao-Yang, WANG Yu-Dong, XIONG Hao, YANG Feng, DONG Chun-Hui, WANG Zhong-Hai, ZHOU Rong, YANG Chao-Wen. Design of portable α - β - γ spectrometer [J]. J Sichuan Univ: Nat Sci Ed, 2020, 57: 947.